Skip to main content
. 2015 May 8;16(3):60–79. doi: 10.1120/jacmp.v16i3.4636

Table 2.

Analysis of systematic errors and statistical dispersions in water for a field of 4×4cm2 at 18 MeV and SSD=100cm. The statistical accuracy was 2% and the grid size was 2×2mm2.

Curve Region Maximum Systematic Error Maximum Statistical Dispersion
dmax – d80% 0% 1.5%
d80% – d20% 0.4 mm 1.5 mm
d20% – end 0.4% 0.8%