Table 7.
Maximum deviation (Δmax) between measured and calculated depth‐dose curves for different combinations of energy/applicator in water phantom at . The max evaluation is separated into four groups: from 0.5 cm to depth to dmax, from dmax to depth of 80%, from depth of 80% to depth of 20%, and from depth of 20% to the end. Deviations are given in term of % dose difference, except in the gradient region (80%‐20%), where it is represented in term of distance to agreement (mm).
Energy (MeV) | Field Size (cm2) |
|
||||
---|---|---|---|---|---|---|
|
–80% | 80%–20% | 20%–end | |||
6 |
|
2.0% | 2.0% | 0.5 mm | 2.0% | |
|
2.0% | 2.0% | 0.3 mm | 2.0% | ||
|
2.5% | 2.5% | 0.5 mm | 2.0% | ||
12 |
|
1.0% | 2.0% | 1.0 mm | 2.5% | |
|
0.3% | 2.5% | 1.0 mm | 1.0% | ||
|
1.0% | 2.0% | 0.5 mm | 1.5% | ||
18 |
|
1.5% | 1.5% | 0.5 mm | 0.3% | |
|
2.0% | 1.5% | 1.0 mm | 0% | ||
|
2.0% | 1.0% | 0.5 mm | 0.3% |