Maximum deviation (Δmax) between measured and calculated depth‐dose curves for different combinations of energy/applicator in water phantom at SSD=100cm. The Δmax evaluation is separated into four groups: from 0.5 cm to depth to dmax, from dmax to depth of 80%, from depth of 80% to depth of 20%, and from depth of 20% to the end. Deviations are given in term of % dose difference, except in the gradient region (80%‐20%), where it is represented in term of distance to agreement (mm).

Energy (MeV) Field Size (cm2)
Δ
max
0.5cmdmax
dmax–80% 80%–20% 20%–end
6
6×6
2.0% 2.0% 0.5 mm 2.0%
10×10
2.0% 2.0% 0.3 mm 2.0%
20×20
2.5% 2.5% 0.5 mm 2.0%
12
6×6
1.0% 2.0% 1.0 mm 2.5%
10×10
0.3% 2.5% 1.0 mm 1.0%
20×20
1.0% 2.0% 0.5 mm 1.5%
18
6×6
1.5% 1.5% 0.5 mm 0.3%
10×10
2.0% 1.5% 1.0 mm 0%
20×20
2.0% 1.0% 0.5 mm 0.3%