Table 3.
Test | Comparison | Description | Tolerance |
---|---|---|---|
5.1 | Dose distributions in planning module vs. modeling (physics) module | Comparison of dose distribution for large () field. | Identical a |
5.2 | Dose in test plan vs. clinical calibration condition b | Reference calibration condition check | 0.5% |
5.3 | Dose distribution calculated in planning system vs. commissioning data | PDD and off axis output factors for a large and a small field size | 2% |
Identical to within the expected statistical uncertainty (considering noise and calculation grid size).
TPS absolute dose at reference point.