TPS model comparison tests and tolerances

Test Comparison Description Tolerance
5.1 Dose distributions in planning module vs. modeling (physics) module Comparison of dose distribution for large (>30×30 cm2) field. Identical a
5.2 Dose in test plan vs. clinical calibration condition b Reference calibration condition check 0.5%
5.3 Dose distribution calculated in planning system vs. commissioning data PDD and off axis output factors for a large and a small field size 2%
a

Identical to within the expected statistical uncertainty (considering noise and calculation grid size).

b

TPS absolute dose at reference point.