Table 5.

Basic TPS photon beam evaluation methods and tolerances

Region Evaluation Method Tolerance a (consistent with IROC Houston)
High dose Relative dose with one parameter change from reference conditions 2%
Relative dose with multiple parameter changes b 5%
Penumbra Distance to agreement 3 mm
Low‐dose tail Up to 5 cm from field edge 3% of maximum field dose
a

Tolerances are relative to local dose unless otherwise noted.

b

For example, off‐axis with physical wedge.