Table 7.
Test | Objective | Description (example) | Detector | Ref |
---|---|---|---|---|
7.1 | Verify small field PDD | MLC shaped field, with PDD acquired at a clinically relevant SSD | Diode or plastic scintillator | Yunice et al.(16) |
7.2 | Verify output for small MLC‐defined fields | Use small square and rectangular MLC‐defined segments, measuring output at a clinically relevant depth for each a | Diode, plastic scintillator, minichamber or microion chamber | Cadman et al. (58) |
7.3 | TG‐119 tests | Plan, measure, and compare planning and QA results to the TG119 report for both the Head and Neck and C‐shape cases | Ion chamber, film and/or array | TG‐119 (Ezzell et al. (37) ) |
7.4 | Clinical tests | Choose at least 2 relevant clinical cases; plan, measure, and perform an in‐depth analysis of the results | Ion chamber, film and/or array | Nelms et al. (42) |
7.5 | External review | Simulate, plan, and treat an anthropomorphic phantom with embedded dosimeters. | Various options exist b | Kry et al. (39) |
A bar pattern scanned with a diode can be used to obtain additional absolute dose profile comparison in the direction perpendicular to MLC movement
If IROC Houston service is used, they typically employ TLDs and radiochromic film. Certain commercial phantoms can accommodate ion chambers for point dose measurements