. 2015 Sep 8;16(5):159–166. doi: 10.1120/jacmp.v16i5.5372

Table 1.

Estimates of effective and equivalent dose and limits for radiation workers (RWs) and the general public. The maximum dose rate measured was above the patient's abdomen (0.4 μSv/hr), and the maximum measured total dose to any staff was 1 μSv/hr

Dose at Distance from Brachytherapy Implant Area Expected Dose Rate (μSv/hr) Expected Dose (mSv)
Staff @ 1 meter with no patient shielding
9.5a
0.04a
Surgeons @ 30 cm 0.12
hands without Pb gloves @ 10 cm 1.12
@ 1 cm 112.3
Surgeons @ 30 cm 0.04
hands with Pb gloves @ 10 cm 0.39
@ 1 cm 39.3
a

The annual equivalent dose limit is 50 mSv in Canada and the United States, or approximately 25 μSv/hr over a 40‐hour work week. The expected dose calculation assumes a 2.5‐hour surgery duration.