Skip to main content
. 2016 Jan 8;17(1):34–48. doi: 10.1120/jacmp.v17i1.5452

Table 2.

Pass rates (mean ± standard deviation) for various gamma criteria for each error in TG‐119 film planes.

Type of Error 3%/3 mm Absolute Gamma Analysis 2%/2 mm Absolute Gamma Analysis 3%/3 mm Relative Gamma Analysis
Incorrect FF material
100%±0%
100%±0%
100%±0%
TMR vs. PDD
53.37%±6.10%
44.97%±4.19%
98.13%±2.34%
Smoothed for Farmer chamber
100%±0%
100%±0%
100%±0%
Electron Contamination
100%±0%
100%±0%
100%±0%
MLC DLG .5 mm
98.03%±1.96%
96.17%±1.96%
100%±0%
MLC DLG 1 mm
77.43%±7.97%
68.4%±6.84%
93.07%±8.22%
Transmission Factor 0.5% change
95.89%±3.55%
95.05%±4.05%
97.33%±2.29%
Transmission Factor 0.9% change
92.02%±5.16%
88.72%±6.49%
95.15%±3.92%