Table 1.

Measurements taken to characterize the EPID and used to input into software to create the model.

Measurement Comment Equipment Phantom (cm×cm×cm) Field Size (cm2)
1. Calibrate ion chamber For absolute dosimetry and to account for daily output variation Farmer chamber used to calibrate Semiflex back to absolute dose 30×30×20 slabs
10×10
2. Sij matrix (and offset) To measure dose response over entire EPID. a) Semiflex with a brass buildup in an empty PTW MP3 water tank b) Take 2 EPID images
26×26
3. Open‐field series At 160 cm SSD a) Semiflex with brass buildup b) EPID Series
4. Phantom field series Constant phantom thickness, varying field size a) Semiflex at isocenter of phantom b) EPID 30×30×20 slabs Series
5. Phantom thickness series Constant field size, varying phantom thickness a) Semiflex at isocenter of phantom b) EPID Series
10×10