Skip to main content
. 2016 Jan 8;17(1):112–120. doi: 10.1120/jacmp.v17i1.5686

Table 2.

The difference in dose between the planned and measured doses are shown for different field sizes and phantom thicknesses. The dose differences (Δdose (%)) are show for flattened and FFF beams.

4cm×4cm
10cm×10cm
20cm×20cm
Field Size 6 MV 6 MV FFF 6 MV 6 MV FFF 6 MV 6 MV FFF
12 cm ‐3.5 +0.2 +0.1 ‐0.1 +1.0 +2.1
24 cm ‐0.4 +0.8 +0.1 +0.1 ‐0.6 ‐0.6
36 cm ‐2.5 +0.9 ‐0.3 +0.4 ‐3.1 ‐5.2