Table 4.
IMRT FFF SABR lung results showing the difference in dose between the planned and measured doses for the different fields. Also shown are the percentage of points with a gamma value less than or equal to 1, using a global gamma of 3%/3 mm and 5%/3 mm. Measurements taken with a Delta4 are also included.
1 (145°) | 2 (196°) | 3 (247°) | 4 (298°) | 5 (349°) | 6 (40°) | 7 (91°) | Total | ||
---|---|---|---|---|---|---|---|---|---|
|
‐8.7 | +2.1 | +2.8 | ‐1.8 | ‐2.0 | +1.2 | +0.4 | ‐0.6 | |
|
100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | |
|
87.6 | 97.8 | 100.0 | 100.0 | 89.3 | 87.0 | 92.7 | 89.5 | |
|
89.8 | 98.0 | 100 | 100 | 90.3 | 88.8 | 94.7 | 99.3 |