Table 1. The dosimetric variation of the nominal plan and RR plan under scenarios of rotated patient planning image CT (-3 and +3 degree along the superior-inferior axis) and verification CT scan (VFCT1 and VFCT2). The is defined as the percentage change of each scenario relative to the plan on the planning CT, which is less affected by the slight dose difference between the nominal plan and RR plan and helps to compare the robustness of the two plans.
iGTV D95 | iGTV D50 | CTV D95 | CTV D50 | |
The nominal plan | ||||
-3 degree (CCW) | -10.1% | -4.0% | -14.5% | -3.8% |
+3 degree (CW) | -5.9% | -1.8% | -5.8% | -2.1% |
VFCT1 | -20.2% | -8.9% | -25.6% | -9.9% |
VFCT2 | -9.2% | -2.0% | -13.8% | -1.8% |
RR plan | ||||
-3 degree (CCW) | -1.2% | -0.8% | -1.2% | -0.2% |
+3 degree (CW) | -3.3% | -1.4% | -3.2% | -1.0% |
VFCT1 | -5.4% | -2.4% | -7.3% | -1.8% |
VFCT2 | -2.7% | -0.3% | -4.7% | 0.0% |