Table 1.

Gamma passing rates (3% and 3 mm) for EBT‐XD and EBT3 films

Central Axis Dose (cGy) EBT‐XD Red Channel EBT3 Red Channel EBT3 Green Channel
Absolute Relative Absolute Relative Absolute Relative
Plan 1 1,277 100.0% 100.0% 91.6% 99.7% 98.9% 98.7%
Plan 2 1,302 99.1% 99.1% 80.9% 99.9% 96.2% 99.4%
Plan 3 1,704 98.9% 100.0% 100.0% 100.0% 98.9% 99.7%
Plan 4 1,266 98.9% 99.9% 83.8% 98.0% 92.8% 97.8%

Absolute=absolute dose; Relative=relative dose normalized to the isocenter.