. 2016 May 8;17(3):111–127. doi: 10.1120/jacmp.v17i3.5959

Table 1.

Positioning uncertainty model for the SRT/SRS treatment with the frameless ExacTrac Brainlab system.

ΔM
Mask systematic uncertainty Mean of shifts from the XC data for each patient
σS
Setup random uncertainty Standard deviation of the XC data for each patients
ΔF
IR frame systematic uncertainty Mean of the XC data for all patients
ΔL
Localizer systematic uncertainty Mean of XV data
σXR
X‐ray image quality and image registration algorithm Standard deviation of means of XV data
ΔC
Couch sagging Phantom study with various couch angles and weights
ΔISOMV
Radiation isocenter Machine Winston‐Lutz QA data
ΔISOkV
Imaging isocenter Machine Winston‐Lutz QA data
ΔISOIR
IR optical imaging isocenter Machine Winston‐Lutz QA data