. 2016 May 8;17(3):111–127. doi: 10.1120/jacmp.v17i3.5959

Table 4.

Mean and standard deviation (σ) of the position offset for the gantry, collimator, couch rotations for the MV machine, and kV gantry rotation using Winston‐Lutz test.

Isocenter Measurements
(Mean±σ)
Lat (mm) Long (mm) Vert (mm)
Gantry Radiation Isocenter
0.27±0.27
0.24±0.24
0.34±0.34
Collimator Radiation Isocenter
0.32±0.21
0.24±0.14
n/a
Couch Radiation Isocenter
0.32±0.21
0.22±0.13
n/a
X‐ray
0.15±0.80
0.40±0.16
0.21±0.09
Infrared Isocenter
0.02±0.01
0.02±0.01
0.02±0.01