. 2016 May 8;17(3):111–127. doi: 10.1120/jacmp.v17i3.5959

Table 6.

Summary of the individual and cumulative systematic and random uncertainties for the different components of the SRS/SRT treatment with the Brainlab ExacTrac system.

Uncer. SRS/SRT Equipment Type Lat (mm) Long (mm) Vert (mm) Roll (o) Pitch (o) Yaw (o) Rad (mm)
ΔM
Mask Systematic Depends on the mask manufactured for each patient as shown in Fig. 1.
σS
Setup Random Standard Deviation of Mean Quad Sum XC
ΔF
Frame Systematic 0.18
0.25p=0.01
1.27p=0.17
0.32p=0.33
0.18p=0.59
0.47p=0.60
2.52
ΔL
Localizer Systematic
0.03p=0.33
0.01
0.03
0.03
0.00
0.01
0.36
σXR
X‐ray Random 0.20 0.18 0.16 0.26 0.20 0.19 0.16
ΣISOMV
Isocenter of MV beam Systematic 0.27 0.24 0.34 n/a n/a n/a
ΣISOKV
Isocenter of kV imager Systematic 0.15
0.40
0.21 n/a n/a n/a
ΣISOIR
Isocenter of IR system Systematic 0.02 0.02 0.02 0.01 0.01 0.01
ΔC
Couch sag Systematic 0.85 0.01
0.68
0.10
0.20 n/a 1.09