Routine kV‐CBCT QA test items and tolerance used at our institution.

Test Items Daily Monthly Tolerance
Safety: collision interlocks X X Functional
Rotation center and machine isocenter coincidence X X 1.5 mm
CBCT image registration accuracy X 1 mm
Couch movement accuracy after registration X 1 mm
kV image and machine isocenter coincidence X 1 mm
2D kV image quality: low contrast resolution X 12 disks
2D kV image quality: high contrast resolution X 10 line‐pair
CBCT image quality: uniformity and noise X 2%/2%
CBCT image quality: low contrast resolution X 2%
CBCT image quality: high spatial resolution X 4 line‐pair a
CBCT image quality: geometrical accuracy X 1.04 mm
a

The baseline value at the time of installation was 7 line‐pair (lp)/cm, but later the tolerance level was set to a realistically achievable value of 4 lp/cm.