Skip to main content
. 2016 May 8;17(3):211–222. doi: 10.1120/jacmp.v17i3.6047

Table 2.

Imaging parameters used for routine QA and clinical imaging on the XVI system on Elekta Synergy.

Protocol QA Head&Neck Chest Pelvis
Voltage (kVp) 120 100 120 120
Voxel size (mm)
0.5×0.5×2.5
1×1×2
1×1×3
1×1×3
Exposure (mAs) 1056 36.6 650 1056
Rotation angle 360 200 360 360
Filter types S20, F0 S20, F0 M20, F0 M20, F0