Table 1.

Evaluation of R100% in SBRT lung plans calculated by iPlan XVMC algorithm. First 5 patients had dose delivery schema of total 54 Gy in 3 fractions, and remaining 15 patients with 50 Gy in 5 fractions.

Patient No. PTV Vol. (cc) RTOG 0915 Minor Deviation R100% XVMC
1 11.1 1.2‐1.5 1.19
2 20.1 1.2‐1.5 1.32 a
3 27.1 1.2‐1.5 1.20
4 50.9 1.2‐1.5 1.13
5 68.1 1.2‐1.5 1.31 a
6 12.9 1.2‐1.5 1.08
7 15.0 1.2‐1.5 1.34 a
8 20.1 1.2‐1.5 1.31 a
9 21.0 1.2‐1.5 1.28 a
10 26.3 1.2‐1.5 1.20
11 31.2 1.2‐1.5 1.20
12 36.2 1.2‐1.5 1.42 a
13 37.1 1.2‐1.5 1.32 a
14 39.7 1.2‐1.5 1.25 a
15 39.8 1.2‐1.5 1.12
16 53.4 1.2‐1.5 1.30 a
17 56.5 1.2‐1.5 1.37 a
18 61.5 1.2‐1.5 1.19
19 132.0 1.2‐1.5 1.20
20 163.0 1.2‐1.5 1.17
AVG 46.1 1.25
STDEV 38.7 0.09
a

Data that have minor deviations from RTOG 0915 criteria.

PTV = planning target volume; R100% = ratio of prescription isodose volume to PTV; AVG = average; STDEV = standard deviation.