Table 4.

Evaluation of normal lung V20 in SBRT lung plans calculated by iPlan XVMC algorithm. First 5 patients had dose delivery schema of total 54 Gy in 3 fractions, and remaining 15 patients with 50 Gy in 5 fractions.

Patient No. PTV Vol. (cc) RTOG 0915 Minor Deviation V20 (%) a XVMC
1 11.1 10‐15 3.6
2 20.1 10‐15 2.2
3 27.1 10‐15 4.2
4 50.9 10‐15 10.0
5 68.1 10‐15 9.9
6 12.9 10‐15 1.5
7 15.0 10‐15 1.8
8 20.1 10‐15 2.1
9 21.0 10‐15 1.6
10 26.3 10‐15 3.1
11 31.2 10‐15 3.3
12 36.2 10‐15 2.3
13 37.1 10‐15 3.3
14 39.7 10‐15 9.2
15 39.8 10‐15 1.1
16 53.4 10‐15 4.7
17 56.5 10‐15 7.4
18 61.5 10‐15 5.7
19 132.0 10‐15 4.3
20 163.0 10‐15 12.8 b
AVG 46.1 4.7
STDEV 38.7 3.4
a

Normal lung V20 values ranged from 1.1% to 12.8% (mean=4.7%±3.4%).

b

Data that have minor deviations from RTOG 0915 criteria. No minor deviation in normal lung V20 from the RTOG 0915 criteria was observed; except for one patient (Patient #20), whose PTV was about 163 cc (the largest PTV recommended by RTOG protocol), had V20=12.8%.

PTV = planning target volume; V20 = percentage of total normal lung minus ITV receiving dose equals larger 20 Gy or more; AVG = average; STDEV = standard deviation.