Table 1.

A comparison between the simulated (RDF) of 103Pd source in this work with the published data by other investigators.

Radial Dose Function, g(r)
Distance from source (cm) This work Thomson et al. (3) TG‐43 (26)
0.05 0.251 0.249 —–
0.06 0.414 0.401 —–
0.07 0.580 0.561 —–
0.08 0.735 0.704 —–
0.09 0.864 0.826 —–
0.1 0.962 0.929 0.911
0.15 1.263 1.226 1.21
0.2 1.363 1.34 —–
0.25 1.383 1.381 1.37
0.3 1.381 1.39 1.38
0.4 1.332 1.365 1.36
0.5 1.292 1.31 1.3
0.6 1.213 1.245 —–
0.7 1.183 1.18 —–
0.75 1.149 1.148 1.15
0.8 1.106 1.113 —–
0.9 1.042 1.059 —–
1 1.001 1.001 1
1.5 0.763 0.742 0.749
2 0.576 0.552 0.555
2.5 0.426 0.407 0.41
3 0.315 0.298 0.302
3.5 0.231 0.219 0.223
4 0.170 0.16 0.163
4.5 0.124 0.117 —–
5 0.092 0.0865 0.0887
5.5 0.067 0.0635 —–
6 0.050 0.0469 0.0482
6.5 0.037 0.0346 —–
7 0.027 0.0256 0.0262
7.5 0.021 0.0193 —–
8 0.015 0.0147 —–
8.5 0.012 0.0112 —–
9 0.009 0.0084 —–
9.5 0.007 0.0064 —–
10 0.005 0.0051 0.00615