Table 2.

Observed dose difference causes and established course of action.

Type of Problem Problem Solution
Patient deform ation & set up Anatomical variation (weight, oedema, gas, etc..…) Verify SSD, thicknesses and replan plan with new CT data if necessary
Unexpected object in the beam Patient setup Review patient setup and replan acquisition
Treatment plan & choice of points of interest Point in low‐dose area
Find new point of interest and recalculate transit dose
Point in high‐dose gradient
Point in high‐dose gradient
Point close to tongue‐and‐groove area
Point close to an heterogeneity interface
Machine & EPID Acquisition interrupted before end of beam irradiation Replan acquisition
EPID in wrong position Calibrate EPID positioning and replan acquisition
Drift of EPID calibration in dose or machine output Proceed to a new calibration