. 2016 May 8;17(3):380–391. doi: 10.1120/jacmp.v17i3.5724

Table 5.

The total effective dose contribution from the given organ for the chest region simulated for the large bowtie filter. This is relative to the total effective dose from a procedure where ODM is not enabled with identical CTDIvol and where TCM is set at the same setting. Only organs which contribute over 5% to the effective dose are shown. The values in brackets indicate the dose change from enabling ODM.

Normalized Relative Effective Dose Contribution
Organ ODM on TCM on ODM off TCM on ODM on TCM off ODM off TCM off
Effective dose (ICRP103) (mSv)
94.7%(5.3%)
100.0%
94.5%(+5.5%)
100.0%
Lungs (mGy)
33.1%(+0.3%)
32.8%
32.8%(+0.2%)
32.6%
Breasts (mGy)
22.4%(5.0%)
27.4%
23.1%(5.3%)
28.4%
Active bone marrow (mGy)
11.2%(+0.6%)
10.6%
11.2%(+0.6%)
10.6%
Oesophagus (mGy)
7.3%(+0.3%)
7.0%
7.3%(+0.3%)
7.0%
Thyroid (mGy)
5.0%(0.5%)
5.5%
5.2%(0.6%)
5.8%