. 2015 Nov 8;16(6):435–448. doi: 10.1120/jacmp.v16i6.5616

Table 1.

Settings of the collimator jaw position according to the applicator and the nominal electron beam energy. All field sizes are defined at SSD=100cm. R100,R50, and RP values are specific for each electron beam energy

Applicator Field Size
6×6
a
(cm2) 10×10
20×20
Energy (MeV) R100 (cm) R50 (cm) RP (cm) Collimator Jaw Settings (cm2)
Siemens Primus KD2
6 1.40 2.45 3.05
16×16
19×19
27×27
9 2.10 3.55 4.36
16×16
19×19
27×27
12 2.90 4.79 5.83
16×16
19×19
27×27
18 3.50 7.57 9.12
16×16
19×19
27×27
Varian 2300C/D
6 1.34 2.37 3.00
20×20
20×20
25×25
9 2.15 3.60 4.39
20×20
20×20
25×25
12 2.88 5.02 6.01
11×11
14×14
25×25
18 3.45 7.64 9.10
11×11
14×14
22×22
Siemens Oncor
6 1.40 2.47 3.16
13×13
19×19
30×30
9 2.10 3.60 4.50
13×13
19×19
29×29
12 2.80 4.82 5.80
13×13
19×19
28×28
14 3.10 5.21 6.30
13×13
19×19
27×27
a

For Siemens Oncor, there is no 6cm×6cm applicator but a circular applicator with 5 cm diameter.