For each of the cycle lengths at which the simulated cylindrical target was driven, the resulting images were segmented and the centroid positions plotted against frame number were fitted to a sinusoidal curve. The fitting parameters from the MATLAB sinusoidal curve are presented here. The target was driven with a 20 mm amplitude at 3–9 s per breath

Cycle Length (s per breath) Amplitude (mm) Measured Angular Frequency (radians/s) Sum of Squared Residuals R‐square Adjusted R‐square Root‐mean‐square Error
3 20.26 0.516 1.99 0.9999 0.9999 0.176
4 20.36 0.389 1.21 0.9999 0.9999 0.131
5 20.37 0.311 1.22 0.9999 0.9999 0.137
6 20.42 0.259 1.85 0.9999 0.9999 0.169
7 20.17 0.225 46.1 0.9966 0.9964 0.848
8 20.35 0.195 1.70 0.9999 0.9999 0.163
9 20.45 0.173 1.37 0.9999 0.9999 0.147
Average 20.34