Water‐equivalent depth of 50% maximum distal dose

Energy (MeV) Modulation (mm) Diode (mm) Markus (mm) Simulation (mm) Diode – Markus (mm) Diode – Geant4 (mm)
127 None 100.8 99.9 99.5 0.9 1.2
127 15 100.4 99.5 99.8 0.9 0.7
127 30 100.5 99.6 99.8 0.9 0.7
157 None 154.2 153.7 153.1 0.5 1.2
157 30 153.6 153.2 153.0 0.4 0.6
157 60 153.8 a 153.0 a 0.8
a

Data unavailable.