Published in final edited form as: Clin Chim Acta. 2017 Mar 10;469:31–36. doi: 10.1016/j.cca.2017.03.010

Table 2.

Mean bias (%) of laboratories’ results against RMP-defined targeta.

Sample (target value)
Assay Manufacturer Analytical System
Sample A (15.5 ng/dl) Sample B (30.0 ng/dl) Sample C (402 ng/dl) Sample D (498 ng/dl) Assay Manufacturer Calibration bias





nb Mean bias in % (95% CI) n Mean bias in % (95% CI) n Mean bias in % (95% CI) n Mean bias in % (95% CI) n Mean bias in % (95% CI)
Beckman Coulter 20 13.1 (8.8 to 17.4)
 Access 2 6 73.1 (58.4 to 87.9) 6 18.7 (10.1 to 27.4) 6 −8.8 (−11.2 to −6.3) 6 −3.9 (−7.5 to −0.4)
 Unicel DxI 800 9 51.7 (32.9 to 70.4) 11 15.8 (9.6 to 21.9) 11 −15.1 (−16.9 to −13.4) 10 −13.2 (−17.0 to −9.5)
Siemens 75 −1.2 (−4.5 to 2.2)
 ADVIA Centaur (XP, XPT) 34 37.9 (26.2 to 49.6) 39 21.3 (13.6 to 29.1) 39 −16.6 (−19.6 to −13.7) 39 −13.3 (−15.8 to −10.8)
 Immulite 1000 4 13.5 (−7.7 to 34.7) 4 −17.6 (−26.8 to −8.4)
 Immulite 2000 XPi 20 9.8 (−1.2 to 20.7) 20 −24.8 (−28.3 to −21.4) 21 −22.1 (−25.6 to −18.6)
 Coat-A-Count 4 −2.6 (−9.5 to 4.3)
Roche Diagnostics 26 3.5 (0.1 to 6.9)
 Cobas e401,e601,e602 5 −11.7 (−30.2 to 6.8) 7 11.1 (−2.4 to 24.5) 7 3.2 (−1.7 to 8.1) 5 2.7 (−4.1 to 9.6)
 MODULAR series 9 −17.8 (−29.7 to −5.8) 11 18.3 (11.8 to 24.7) 11 2.5 (0.7 to 4.3) 10 1.5 (−2.0 to 5.0)
 Elecsys 8 −2.1 (−11.5 to 7.3) 7 18.7 (7.5 to 29.9) 7 8.6 (3.4 to 13.7) 8 4.3 (1.6 to 6.9)
LC-MS/MS 5 25.5 (0.0 to 51.1) 5 3.1 (−2.5 to 8.6) 5 3.9 (−0.2 to 8.1) 5 6.8 (−2.7 to 16.3) 6 8.0 (0.5 to 15.4)
Ortho Clinical Diagnostics 8 −24.1 (−28.6 to −19.6)
a

Results from groups with 4 or more participants are reported.

b

n = number of participating laboratories.