Published in final edited form as: Anal Biochem. 2017 Oct 18;539:81–89. doi: 10.1016/j.ab.2017.10.014

Table 5.

Limit of detection (LOD) and limit of quantification (LOQ) for hCE1 using partially purified enzyme. Substrate activities ([S] = 50 µM) were measured at 37°C in sodium phosphate buffer (100 mM, pH = 8, 0.1 mg/mL BSA) containing 1% DMSO. The signal-to-noise ratio was determined by taking the ratio of the mean signal to the standard deviation of the background signal, each performed as quadruplicates of a single experiment.

Kinetic
Modea
Endpoint mode

1 hour 4 hours 24 hours

4-MOMMP
  LODb 3.9 ng 0.06 ng 0.03 ng 0.008 ng
  LOQc 16 ng 2.0 ng 0.12 ng 0.03 ng
CMNA
  LODb 2.0 ng 0.98 ng 3.9 ng 0.06 ng
  LOQc 3.9 ng 7.8 ng (−)d (−)d
a

Kinetic measurements were taken at 30 second intervals for 30 minutes.

b

LOD is the lowest measurement where S/N > 3.

c

LOQ is the lowest measurement where S/N > 9.

d

The (−) indicates that there were insufficient data points above S/N > 9 to generate a linear curve.