Table 4.
Troubleshooting
PROBLEM | STEP(S) | POSSIBLE REASON | SOLUTION |
---|---|---|---|
Poor cell recovery or high debris | 6, 17, 31, 103 + 104–110 | Low viability of starting cells | Improve cell culture and preparation. |
Incorrect spin speeds | Check spin speeds (note increased speed after fixation) in all centrifugations, particularly Step 51 onwards. | ||
Low starting cell number | Increase cell starting number. | ||
Harsh cell treatment | Limit vortexing of samples and handle carefully throughout the protocol, see Step 35 for correct procedure. | ||
Poor mRNA staining | 103 + 104–110 | Oven temperature incorrect in Steps 76, 78, 85, 91 & 98. | Recalibrate oven. |
Oven temperature unstable in Steps 76, 78, 85, 91 & 98. | Carefully monitor oven temperature throughout protocol | ||
Incorrect amount of mRNA Target Probes used in Step 68. | Use Target Probes at 1:20 | ||
Incorrect amount of Label Probes used in Step 96. | Use Label Probes at 1:100 | ||
High residual volume in tubes, following washes in Steps 73, 81, 89 & 95. | Maintain residual volume at 100 µl. | ||
mRNA degradation in Steps 1–64. See Step 27 for example. | Ensure cells are in the growth phase before use. Work at 4 °C before fixation. | ||
Poor cell permeabilization and/or fixation in Steps 48, 53 & 65. | Check Reagent Setup (see Table 1). Use buffers prepared on the same day. Check incubation times and temperatures. | ||
Poor washing in Steps 71–73, 79–81, 87–89, 93–95, 99–101 | Ensure all wash steps are followed, using the appropriate buffer at the required temperature. | ||
High background in mRNA channel/high MFI of mRNA negative population | 103 +104–110 | Low viability/number of starting cells in Steps 1–24 and 36–40. Check viability in Step 31. | Improve cell culture and preparation. Use Viability stain to exclude dead cells. |
Incorrect amount of mRNA Target Probes used in Step 68. | Use Target Probes at 1:20 | ||
Incorrect amount of Target Probes used in Step 96. | Use at 1:100 | ||
Unclean flow machine/samples run too fast. | Clean the flow cytometer well before use. Run samples at ~2,000 events/second. | ||
Low residual volume following washes in Steps 73, 81, 89 & 95. | Residual volume should be 100 µl. | ||
Incorrect fixation time in Steps 50 + 76. | Check fixation time – longer fixes can increase background. | ||
Poor washing in Steps 71–73, 79–81, 87–89, 93–95, 99–101. | Ensure all wash steps are followed, using the appropriate buffer at the required temperature. | ||
No HIVRNA+/Gag+cells detected | 103 + 104–110 | Low starting cell number in Step 23. | See “Limitations” for discussion of starting cell number. |
Poor mRNA staining. | Run a positive control sample to rule out assay issues. | ||
Poor antibody staining | 103 + 104–110 | Clone unstable during protocol. | Test new antibody clones. See Box 4 for further information. |
Fluorochrome unstable during protocol. | Test additional fluorochromes. See Box 4 for further information. | ||
Incorrect incubation times/temperatures in Steps 43 & 61. | For anti-Gag KC57, stain for 1 hr (30 min at RT, 30 min at 4 °C) | ||
Non-optimal antibody concentration in Step 43. | Titrate all antibodies before use. | ||
Incorrect compensation. | See Box 4 for details | ||
Interactions between BV, BUV and BB polymer dyes. | Prepare antibody mixes with BD Horizon Brilliant Stain Buffer (Cat #563794) and compare to standard mix to determine if this buffer is required. |