Skip to main content
. Author manuscript; available in PMC: 2018 Nov 15.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2017 Oct 5;1068-1069:90–97. doi: 10.1016/j.jchromb.2017.10.008

Table 3.

Accuracy and precision of QC samples (n=3)

Nucleotide Content (fmol)
250 625 2500
dATP
Mean ± SD 236.5 ± 53.5 573.2 ± 27.1 2376.4 ± 138.6
CV (%) 23.0% 5.0% 6.0%
Accuracy (%) 95% 92% 95%
dGTP
Mean ± SD 247.1 ± 59.3 530.4 ± 34.4 2596.4 ± 338.8
CV (%) 24.0% 6.0% 13.0%
Accuracy (%) 99% 85% 104%
dCTP
Mean ± SD 219.9 ± 60.5 530.4 ± 96.4 2465.4 ± 124.4
CV (%) 28.0% 15.0% 5.0%
Accuracy (%) 88% 104% 99%
dTTP
Mean ± SD 230.6 ± 54.2 571.5 ± 115.7 2466.5 ± 116.4
CV (%) 23.0% 20.0% 5.0%
Accuracy (%) 92% 91% 99%