Table 3.
Reproducibility of 2D and 3D MRE stiffness values measured under different breath-hold phases, keeping driver type constant
Conditions | Reproducibility parameter | ||||||
---|---|---|---|---|---|---|---|
|
|
||||||
MRE method |
Breath-hold phase (Variable condition) |
Driver type (Constant condition) |
Bias (kPa)* |
Bias p- value |
RDC | LOA (kPa) |
CCC |
2D | Exp – Ins | Rigid | − 0.08 | 0.14 | 0.44 | −0.52, 0.36 | 0.72 (0.38, 0.89) |
2D | Exp – Ins | Flexible | − 0.07 | 0.28 | 0.54 | −0.62, 0.47 | 0.71 (0.36, 0.88) |
3D | Exp – Ins | Rigid | − 0.19 | <0.01 | 0.31 | −0.50, 0.13 | 0.61 (0.26, 0.81) |
3D | Exp – Ins | Flexible | − 0.13 | <0.01 | 0.30 | −0.42, 0.17 | 0.73 (0.41, 0.89) |
Table summarizes the reproducibility parameters for 2D and 3D MRE measured under different breath-hold phases, keeping driver type constant.
The bias is calculated as the mean value in expiration minus the mean value in inspiration. Hence, a negative bias indicates that liver stiffness values measured in expiration were less than those measured in inspiration.
Ins=inspiration;. Exp=expiration;. RDC=reproducibility coefficients; LOA=Limit of Agreement; CCC=concordance correlation coefficients.
Values in parentheses are 95% confidence limits.