Table 4.
Reproducibility of 2D and 3D MRE stiffness values measured with different driver types, keeping breath-hold phase constant
| Conditions | Reproducibility parameter | ||||||
|---|---|---|---|---|---|---|---|
|
|
|
||||||
| MRE method |
Breath-hold phase (Constant condition) |
Driver type (Variable condition) |
Bias (kPa)* |
Bias p- value |
RDC | LOA (kPa) |
CCC |
| 2D | Inspiration | Rigid – Flexible | 0.07 | 0.22 | 0.44 | −0.37, 0.50 | 0.79 (0.50, 0.92) |
| 2D | Expiration | Rigid – Flexible | 0.06 | 0.20 | 0.37 | −0.31, 0.43 | 0.83 (0.58, 0.94) |
| 3D | Inspiration | Rigid – Flexible | 0.03 | 0.19 | 0.20 | −0.17, 0.23 | 0.94 (0.82, 0.98) |
| 3D | Expiration | Rigid – Flexible | −0.03 | 0.20 | 0.19 | −0.22, 0.16 | 0.86 (0.65, 0.95) |
Table summarizes the reproducibility parameters for 2D and 3D MRE measured with different drivers, keeping breath-hold phase constant.
The bias is calculated as the mean stiffness value with the rigid driver minus the mean value with the flexible driver. Hence, a positive bias indicates that liver stiffness values measured with the rigid driver were greater than those measured with the flexible driver.
RDC=reproducibility coefficients; LOA=Limit of Agreement; CCC=concordance correlation coefficients.
Values in parentheses are 95% confidence limits.