Table 6.

Precision of MGIT PZA susceptibility test results stratified by inoculation method. Reproducibility of categorical results for each duplicate pair of isolates.

Precision Percentage (number) by inoculation method
BD A C
Precisea  & accurate 72.2 (109) 87.6 (127) 93.5 (131)
Precise & inaccurate 15.2 (23) 6.9 (10) 2.9 (4)
Not preciseb 12.6 (19) 5.5 (8) 3.6 (5)

Subtotalc 151 145 140
<2 resultsd 1.3 (2) 5.2 (8) 8.5 (13)

Totale 153 153 153

aEach duplicate isolate had the same categorical susceptibility result. bEach duplicate isolate had a different categorical susceptibility result. cDenominators used to calculate precision percentages. dTime-Out occurred in one or both of the duplicate pairs of isolates. eDenominator used to calculate <2 results percentages.