. 2017 Jul-Sep;10(3):132–135. doi: 10.5935/1984-0063.20170023

Table 1.

Mean ± SD values for the measured sleep variables between non-dominant and dominant wrist-actigraphy devices. Comparison between devices are reported using mean bias, Pearson correlations (r), typical error of estimates (TEE) and coefficient of variation % with 95% confidence intervals.

  Non-Dominant Dominant Mean bias Pearson's r TEE CV%
Total Sleep Time (mins) 436±72 442±80 6±19 0.97 (0.95 - 0.98) 18 (15 -22) 4.6 (3.8 - 4.7)
Total Time in Bed (mins) 552±85 554±83 2±20 0.97 (0.95 - 0.98) 20 (17 - 25) 3.8 (3.1 - 4.7)
Sleep Efficiency (%) 81±9 82±6 1±4 0.89 (0.81 - 0.94) 4 (3 - 5) 5.3 (4.4 - 6.6)
Sleep Latency (mins) 29±34 28±31 -1±15 0.89 (0.82 - 0.94) 15 (13 - 19) 102.4 (80.2 - 140.7)
Wake After Sleep Onset (mins) 53±38 52±43 -1±28 0.76 (0.61 - 0.86) 25 (21 - 32) 60.5 (47.7 - 82.5)
Sleep Onset Time (time of day) 23:25±1:32 23:24±1:33 -1±11 0.99 (0.99 - 0.99) 12 (11 - 15) 0.8 (0.7 - 1.0)