Table 1.

Performance criteria applied to CRMLN lipid reference laboratory using BQ RMP.

Lipid Precision Accuracy
BFC CV ≤ 1.5% ±(CDC LDL-C reference value × 0.02 + HDL-C bias vs. CDC)
[max = ±2 mg/dL or 0.04 (HDL-C reference value) if smaller]
HDL-C SD ≤ 1 mg/dL ±CDC HDL-C reference value × 0.04
LDL-C CV ≤ 1.5% ±CDC LDL-C reference value × 0.02

CRMLN: Cholesterol Reference Method Laboratory Network. BQ RMP: Beta quantification reference measurement procedure.

CDC: US Centers for Disease Control and Prevention.

BFC: Bottom fraction cholesterol.