TABLE 1.

Analysis of IgG concordant and discrepant sera between ICT and Toxo IgG Architect test by testing of 767 nonselected samples

IgG analysis n Architect valuea (UI/ml) ICTb Reference Western blotc (LDBio Toxo II IgG) Conclusion
Concordant 582 0–1.3 (Neg) Neg NR Neg IgG
Concordant 143 ≥3 Pos NR Pos IgG
Minor discrepant 20 1.6–2.9 (gray zone) Pos Pos Pos IgG
Discrepant 14 0.6–1.5 (Neg) Pos Pos Pos IgG (false-negative Architect)
Discrepant 1 2.5 (gray zone) Pos Neg Neg IgG (false-positive ICT)
Discrepant 7 0.1–0.3 (Neg) Pos Neg Neg IgG (false-positive ICT)
a

IgG concentrations on Architect: negative, ≤1.5 IU/ml; gray zone, 1.6 to 2.9 IU/ml; positive, ≥3 IU/ml. Neg, negative.

b

Pos, positive.

c

NR, not realized.