Skip to main content
. Author manuscript; available in PMC: 2017 Nov 28.
Published in final edited form as: Invest Radiol. 2016 Aug;51(8):469–482. doi: 10.1097/RLI.0000000000000256

Table 1.

Sequences and Protocols Used for 3 T and 7 T

T1 MP2RAGE 3D
T2 TSE 2D
FLAIR TSE 2D
T2* GRE
FLAIR SPACE 3D
DWI Single Shot
3 T 7 T 3 T 7 T 3 T 7 T 3 T 7 T 3 T 7 T 3 T 7 T
Voxel dimensions 0.9 × 0.9 0.8 × 0.8 0.5 × 0.5 0.4 × 0.4 0.7 × 0.7 0.7 × 0.7 0.7 × 0.7 0.3 × 0.3 1.1 × 1.0 0.7 × 0.7 1 × 1 1 × 1
Matrix size 228 × 256 300 × 320 345 × 448 388 × 528 240 × 320 240 × 320 240 × 320 520 × 640 194 × 256 320 × 320 222 × 222 222 × 222
No. slices 160 192 23 34 23 34 72 96 160 144 23 34
Field of view, mm2 208 × 230 225 × 240 173 × 230 169 × 230 173 × 230 173 × 230 173 × 230 169 × 208 211 × 250 230 × 230 230 × 230 230 × 230
Slice thickness, mm 0.9 0.8 5 3 5 3 1.7 1.2 1 1 5 3
TE, ms 5 4.13 111 69 127 123 20 15 298 467 90 56
TI, ms 2500 2600
TR, ms 5000 5000 4560 7000 8500 9000 28 27 7000 9000 4200 4500
TA, min:s 8:02 8:02 3:36 3:39 4:00 3:56 5:24 5:16 7:14 7:14 2:05 2:24
GRAPPA factor 2 3 2 2 3 2 3 3 3 3 4
RB/pixel, Hz/pixel 170 180 169 287 150 244 120 140 300 460 1186 1186
FA, degree 4 4 180 180 180 180 15 15 15 15 180 180
Fat saturation Yes Yes

fMRI
T2 SPACE 3D
DWI RESOLVE
T1 FLASH
DIR SPACE 3D
3 T 7 T 3 T 7 T 3 T 7 T 3 T 7 T 3 T 7 T

Voxel dimensions 2.4 × 2.4 1.4 × 1.4 0.7 × 0.7 0.7 × 0.7 1.3 × 1.3 1.2 × 0.9 0.5 × 0.5 0.5 × 0.4 1.3 × 1.3 1.1 × 1.0
Matrix size 96 × 96 160 × 160 288 × 320 288 × 320 180 × 180 185 × 244 336 × 448 315 × 512 192 × 192 191 × 256
No. slices 44 84 128 160 28 34 23 34 144 144
Field of view, mm2 230 × 230 224 × 224 207 × 230 207 × 230 230 × 230 230 × 230 173 × 230 173 × 230 250 × 250 212 × 256
Slice thickness, mm 2.4 1.4 1 0.9 5 3 5 3 1.3 1
TE, ms 26 20 121 87 63 58/97 4.9 3 324 322
TI, ms 3000 3120
TR, ms 3000 3870 2000 2000 4100 6000 253 250 7400 8000
TA, min:s 7:47 7:45 6:06 6:20 4:14 4:32 2:18 2:16 6:12 6:10
GRAPPA factor 3 4 2 3 2 3 2 2 3
RB/pix, Hz/pix 1212 1738 363 446 731 732 170 210 789 592
FA, degrees 90 90 90 90 180 180 80 30
Repetitions 150 120

MP2RAGE indicates magnetization prepared rapid acquisition gradient echo sequence with bias field correction; T2 TSE 2D, 2D turbo spin echo; FLAIR TSE 2D, 2D FLAIR TSE; T2* GRE, T2*-weighted gradient echo sequence; FLAIR SPACE 3D, 3D FLAIR TSE with variable flip angle; fMRI, functional MRI.