Table 8.

Comparison of 7 T/3 T SNRs Calculated From Phantom and In Vivo Estimations

7 T/3 T SNRs
Phantom Volunteers Patients Analyzed as Volunteers Epilepsy Multiple Sclerosis Tumor Vascular Disease
T1 MP2RAGE 3D 0.8 0.8 0.9 0.8 0.7 0.9 0.9
T2 TSE 2D 2.2 2.0 1.8 1.7 2.2 1.7 2.0
FLAIR TSE 2D 1.5 1.1 1.2 1.1 1.2 1.0 1.1
T2* GRE 2.5 3.9 3.6 3.1 3.6 3.8 4.4
f MRI 1.3 1.7
T2 SPACE 3D 1.4 1.3 1.3 1.3 1.4 1.5 1.7
FLAIR SPACE 3D 2.1 2.2 2.3 2.3 2.0 1.9 2.5
DWI single shot 2.5 2.0 1.8 1.9 1.6 1.2
DWI RESOLVE 1.9 1.9 1.7 2.0 2.0 1.4
T1 FLASH 4.2 1.9 1.9 2.3 2.1 1.8 1.8
DIR SPACE 3D 4.2 3.9 3.4 2.9 2.7 2.9 2.8

SNR indicates signal-to-noise; MP2RAGE, magnetization prepared rapid acquisition gradient echo sequence with bias field correction; T2 TSE 2D, 2D turbo spin echo; FLAIR TSE 2D, 2D FLAIR TSE; T2* GRE, T2*-weighted gradient echo sequence; f MRI, functional MRI; T2 SPACE 3D, 3D T2-weighted TSE with variable flip angle; FLAIR SPACE 3D, 3D FLAIR TSE with variable flip angle; DWI single shot, single-shot diffusion-weighted imaging EPI; DWI RESOLVE, multishot readout-segmented diffusion-weighted EPI with 2D navigator correction; T1 FLASH, T1-weighted spoiled gradient echo sequence; DIR SPACE 3D, 3D double inversion recovery TSE with variable flip angle.