Table 4.

Recommended channels for cardiorespiratory polysomnography. The table presents the functions to be assessed, the corresponding biosignal, the technical requirements, and the technical specifications in terms of optimal sampling rate and filter settings

Function Parameter Technical requirements Optimal sampling rate Filter
Sleep EEG, EOG Electrodes 500 Hz 0.3–35 Hz
EMG Electrodes 500 Hz 10–100 Hz
Breathing Oronasal airflow Nasal pressure cannula, thermistor 100 Hz 0.1–15 Hz
Respiratory effort Induction plethysmography 100 Hz 0.1–15 Hz
Oxygen saturation SaO2 25 Hz
Carbon dioxide saturation tcPaCO2 25 Hz
Snoring Microphone 500 Hz
Cardiac ECG Electrodes 500 Hz 0.3–70 Hz
Movement EMG tibialis muscle Electrodes 500 Hz 10–100 Hz
Body position Position sensor 1 Hz
Video Video camera 5 Hz

EEG electroencephalogram, EOG electrooculogram, EMG electromyogram, ECG electrocardiogram, SaO 2 oxygen saturation, tcPaCO 2 transcutaneous measurement of carbon dioxide partial pressure