Table 4.
Function | Parameter | Technical requirements | Optimal sampling rate | Filter |
---|---|---|---|---|
Sleep | EEG, EOG | Electrodes | 500 Hz | 0.3–35 Hz |
EMG | Electrodes | 500 Hz | 10–100 Hz | |
Breathing | Oronasal airflow | Nasal pressure cannula, thermistor | 100 Hz | 0.1–15 Hz |
Respiratory effort | Induction plethysmography | 100 Hz | 0.1–15 Hz | |
Oxygen saturation | SaO2 | 25 Hz | – | |
Carbon dioxide saturation | tcPaCO2 | 25 Hz | – | |
Snoring | Microphone | 500 Hz | – | |
Cardiac | ECG | Electrodes | 500 Hz | 0.3–70 Hz |
Movement | EMG tibialis muscle | Electrodes | 500 Hz | 10–100 Hz |
Body position | Position sensor | 1 Hz | – | |
Video | Video camera | 5 Hz | – |
EEG electroencephalogram, EOG electrooculogram, EMG electromyogram, ECG electrocardiogram, SaO 2 oxygen saturation, tcPaCO 2 transcutaneous measurement of carbon dioxide partial pressure