Table 1. Equilibration time for sucrose (RH/RP = 0.00235) in the dialysis SPR chamber (Deff/D0 = 0.990) of different thicknesses, with (theoretical) and without (experimental) stirring a .
D (mm) | t eff stirred (theoretical) (s) | Normalized ΔλSPR | t eff unstirred (experimental) (s) |
0.15 | 18 | 74 ± 7% | 5529 |
0.30 | 35 | 67 ± 5% | 9892 |
0.60 | 64 | 42 ± 2% | 26 590 |
1.00 | 96 | 39 ± 2% | 55 687 |
a R H values obtained from Pappenheimer et al.1