Skip to main content
. Author manuscript; available in PMC: 2017 Dec 1.
Published in final edited form as: Anal Chem. 2016 Oct 27;88(22):11123–11129. doi: 10.1021/acs.analchem.6b03220

Table 3.

Bias Assessment Using JCTLM-Recognized Reference Materials

IRMM BCR material reference target value, pM (pg/mL) bias, % (95% CI)
BCR 576 114 (31.1) 0.3 (0.2–0.5)a
BCR 577 690 (188) −0.3 (−2.4–1.8)a
BCR 578 140 (365) −0.9 (−4.0–2.2)a
a

No significant bias when evaluated following the protocol described in NIST Special Publication 829.26