Table 1.

Phases of VHP cycles

Phase (duration)
Trial Conditioning (12 min) G1 gassing (12 to 25 min) G2 gassing (2 h 15 min) Dwell (20 min) Aeration (25 to 45 min) Completion (30 to 45 min)
Vaporizer temperature (°C)
 1 30.0 119.7 120.2 120.3 123.7 28.5
 2 30.3 119.6 120.1 120.1 122.7 26.4
 3 26.3 119.7 119.9 120.2 121.3 27.5
 4 23.4 119.7 120.2 117.1 122.0 28.0
 Mean 27.5 119.7 120.1 119.4 122.4 27.6
VHP concentration (ppm)
 1 148 280 212 10
 2 199 343 248 10
 3 126 304 219 10
 4 127 319 207 10
 Mean 151 315 315 10
Airflow (m3/h)
 1 17.7 43.7 42.6 42.8 44.4 49.3
 2 17.8 43.1 41.7 41.9 42.9 48.2
 3 16.4 44.2 43.4 44.0 44.5 46.1
 4 16.4 44.5 43.6 43.6 45.4 46.2
 Mean 17.1 43.9 42.8 43.1 44.3 47.5
Relative humidity (%)
 1 46.7 50.2 72.7 87.2 79.4 49.1
 2 41.6 46.3 75.0 93.5 84.2 46.6
 3 50.4 52.7 75.0 95.8 86.1 53.4
 4 52.8 53.2 75.0 99.4 89.4 58.4
 Mean 47.9 50.6 74.4 94.0 84.8 51.9
Ambient temperature (°C)
 1 25.6 23.9 24.5 25.3 25.2 26.4
 2 26.6 24.1 24.6 25.2 25.0 26.4
 3 24.5 24.0 24.7 24.5 25.2 25.6
 4 24.1 23.8 24.2 25.3 25.1 25.8
 Mean 25.2 24.0 24.5 25.1 25.1 26.1

As cycle phases completed, the vaporizer temperature emitted increased from the beginning of gassing until the end of aeration. The initial conditioning phase increases the temperature of the VHP vaporizer. Once optimal temperature is generated, the first gassing (G1 phase) is initiated followed by an increase of hydrogen peroxide vapor which is then detected in parts per million (ppm). Mean peak VHP concentration occurred during the dwell phase. Relative humidity and ambient temperature increased and decreased relative to cycle phases.