Table 2.
Static–open VHP exposure (7 cycles, involving a total of 28 IVC racks and 17 AHU) | |||
S xylosus PCR analysis of processing equipment | Soiled | Washed | VHP exposed |
No. positive/no. tested (%) | 23/31 (94%) | 6/25 (24%) | 17/48 (35%)# |
No. of CI with ≥6 log kill/ no. tested | No. of BI with no growth/ no. tested | ||
Room VHP indicators | 21/21 (100%) | 7/7 (100%) | |
Equipment VHP indicators (open end)a | 4/4 (100%) | 3/3 (100%) | |
Equipment VHP Indicators (closed end)b | 1/18 (6%) | 0/8 (0%) | |
Active–closed VHP exposure (5 cycles, involving 20 IVC racks and 11 AHU) | |||
S xylosus PCR analysis of processing equipment | Soiled | Washed | VHP exposed |
No. positive/no. tested (%) | 10/10 (100%) | 2/12 (17%) | 0/12 (0%)d |
No. of CI with ≥6 log kill/ no. tested | No. of BI with no growth/ no. tested | ||
Room VHP indicators | 15/15 (100%) | 5/5 (100%) | |
Equipment VHP indicators (sealed)c | 18/18 (100%) | 3/4 (75%) |
Indicators placed inside an unsealed hose, or the plenum or manifold's open, uncapped end, or the AHU exhaust prefilter chamber left open.
Indicators were placed inside the plenum or manifold's capped end, or the AHU exhaust prefilter chamber left closed.
Indicators were placed inside the sealed and connected plenum, manifold, or AHU exhaust prefilter chamber, with the AHU set to the VHP cycle.
P = 0.008