Table 2. Bland-Altman Analysis of Intraocular Pressure Measurements Comparing Results From GAT and Icare HOME Performed by the Investigator and by Self-tonometry.
GAT vs Icare HOME | LoA (95% CI), mm Hg | Width, mm Hg | Fixed Bias, mm Hg | SE (95% CI) | |
---|---|---|---|---|---|
Lower 95% | Upper 95% | ||||
Participant (n = 165) | −4.71 (−4.14 to −5.28) | 9.88 (9.31 to 10.45) | 14.59 | 2.58 | 0.29 (2.01 to 3.15) |
Investigator (n = 198) | −3.48 (−3.91 to −3.05) | 8.8 (8.37 to 9.23) | 12.28 | 2.66 | 0.22 (2.22 to 3.10) |
Abbreviations: GAT, Goldmann applanation tonometer; LoA, limit of agreement.