. 2017 Oct 12;135(10):1030–1036. doi: 10.1001/jamaophthalmol.2017.3151

Table 2. Bland-Altman Analysis of Intraocular Pressure Measurements Comparing Results From GAT and Icare HOME Performed by the Investigator and by Self-tonometry.

GAT vs Icare HOME LoA (95% CI), mm Hg Width, mm Hg Fixed Bias, mm Hg SE (95% CI)
Lower 95% Upper 95%
Participant (n = 165) −4.71 (−4.14 to −5.28) 9.88 (9.31 to 10.45) 14.59 2.58 0.29 (2.01 to 3.15)
Investigator (n = 198) −3.48 (−3.91 to −3.05) 8.8 (8.37 to 9.23) 12.28 2.66 0.22 (2.22 to 3.10)

Abbreviations: GAT, Goldmann applanation tonometer; LoA, limit of agreement.