Table 1.
Conventional linac workflow/standard margins [mm] |
MR-Linac workflow/reduced margins [mm] |
|||||
---|---|---|---|---|---|---|
LR | SI | AP | LR | SI | AP | |
Contouring error | 2.0 | 2.0 | 2.0 | 1.0 | 1.0 | 1.0 |
Difference in observed 4D tumour motion between planning and fraction | 1.2 | 1.6 | 1.4 | – | – | – |
Total systematic error | 2.3 | 2.6 | 2.4 | 1.0 | 1.0 | 1.0 |
Intrafractional baseline drift | 1.5 | 1.7 | 1.8 | 1.5 | 1.7 | 1.8 |
Per-fraction localization error | 2.0 | 2.0 | 2.0 | 1.0 | 1.0 | 1.0 |
Total random error | 2.5 | 2.6 | 2.7 | 1.8 | 2.0 | 2.1 |
Beam penumbra | 6.4 | 6.4 | 6.4 | 8.0 | 8.0 | 8.0 |
Total margin | 6.4 | 7.1 | 6.8 | 2.8 | 2.9 | 2.9 |