. 2014 May 8;15(3):164–181. doi: 10.1120/jacmp.v15i3.4688

Table 4.

VC‐determined offsets (mm) between the treatment isocenter and electronic graticule (as displayed on the OBI workstation) with IsoCal corrections applied for the five linacs. The data are scaled to the plane of the isocenter in the lateral (X) and longitudinal (Y) directions in the panel coordinate system

Source Angle MI M2 M3 M4 M5
X Y X Y X Y X Y X Y
MV
0.2
0.4
0.3
0.9
0.1
0.7
0.3
0.3
0.2
0.7
90°
0.5
0.4
0.7
0.6
0.5
0.4
0.4
0.2 0.2
0.7
180° 0.3
0.4
0.0
0.7
0.3
0.6
0.1 0.0 0.1
0.7
270° 0.5
0.4
0.6
0.6
0.2
0.4
0.4
0.4
0.4
0.8