. 2014 Nov 8;15(6):81–91. doi: 10.1120/jacmp.v15i6.4877

Table 2.

Dosimetric properties of the ExacTrac® localization system using the Unfors RaySafe Xi R/F.

Left Detector Right Detector
Dose (μGy)
    Cranial Standard
20.4±0.2
20.3±0.3
    Thorax Standard
191.8±0.7
189.7±0.8
    Abdomen Standard
154.2±0.7
151.0±0.7
    Pelvis Standard
192.2±0.6
189.7±0.8
Energy (kVp)
    Cranial Standard
81.6±0.5
81.6±0.5
    Thorax Standard
122.5±0.5
120.8±0.5
    Abdomen Standard
122.0±0.8
120.9±0.6
    Pelvis Standard
122.1±0.7
121.3±0.7

Note: Parameter values are mean values of 36 measurements.