Table 1.

A comparative list of some of the characteristics of the two MLCs. Both the 160 MLC and the 120 MLC are in clinical use in our department.

160 MLC on ARTISTE 120 MLC on 2300 EX
Source collimator distance 460 mm 535 mm
Patient clearance 430 mm 334 mm
Leaf resolution 5 mm leaf entire 40 cm field 5 mm central 40 leaves and 10 mm remaining outer leaf
Leaf End S‐shaped C‐shaped
Leaf height 95 mm 55 mm
Penumbra (80 ‐ 20%) 3.6 mm 3.2 mm
Leakage and transmission 0.37% 2.00%
Inter‐digitization Yes Yes
Method to avoid a direct ray in between leaves Tilted alternating pattern of upper and lower leaf tongue and groove
Max Leaf Speed
40mms1
20mms1
Leaf Span 200 mm 145 mm
Step and shoot IMRT Yes Yes
Dynamic IMRT No Yes