Table 1.
160 MLC on ARTISTE | 120 MLC on 2300 EX | |||
---|---|---|---|---|
Source collimator distance | 460 mm | 535 mm | ||
Patient clearance | 430 mm | 334 mm | ||
Leaf resolution | 5 mm leaf entire 40 cm field | 5 mm central 40 leaves and 10 mm remaining outer leaf | ||
Leaf End | S‐shaped | C‐shaped | ||
Leaf height | 95 mm | 55 mm | ||
Penumbra (80 ‐ 20%) | 3.6 mm | 3.2 mm | ||
Leakage and transmission | 0.37% | 2.00% | ||
Inter‐digitization | Yes | Yes | ||
Method to avoid a direct ray in between leaves | Tilted alternating pattern of upper and lower leaf | tongue and groove | ||
Max Leaf Speed |
|
|
||
Leaf Span | 200 mm | 145 mm | ||
Step and shoot IMRT | Yes | Yes | ||
Dynamic IMRT | No | Yes |