Skip to main content
. 2014 Nov 8;15(6):204–216. doi: 10.1120/jacmp.v15i6.4994

Table 3.

Summary of trajectory log file mechanical delivery errors as a function of treatment site for prostate (Pros), prostate and pelvic node (PPN), and head and neck (H&N). Data are presented as mean ±1SD.

Pros PPN H&N
Average Leaf Errors (mm)
0.001±0.002
0.001±0.002
0.001±0.003
Std Dev. Leaf Errors (mm)
0.134±0.009
0.162±0.009
0.138±0.012
RMS Leaf Errors (mm)
0.128±0.008
0.157±0.009
0.135±0.013
Gantry Angle Errors (deg)
0.025±0.006
0.024±0.005
0.026±0.012
Std Dev. Gantry Angle Errors (deg)
0.051±0.003
0.049±0.001
0.059±0.016