Recommended minimum practices for commissioning and QA of an IGRT system

Acceptance Testing and Commissioning
Procedure
    Customer acceptance procedures
    TPS integration
    OIS integration
    Establish routine QA baselines
    QA documentation
Routine Quality Assurance
Procedure Tolerance
Daily
Safety/interlocks Functional
    Imaging‐treatment isocenter coincidence (SRS only) 1 mm
    Positioning/repositioning (SRS only) 1 mm
    Imaging‐treatment isocenter coincidence (SBRT only) 2 mm
    Positioning/repositioning (SBRT only) 2 mm
Weekly
Imaging‐treatment isocenter coincidence (non‐SRS/SBRT) 2 mm
Positioning/repositioning (non‐SRS/SBRT) 2 mm
Semi‐annually
Image scaling 2 mm
Annually
Imaging dose
    2D MV ±1cGy of baseline value
    2D kV (static imaging mode) ±3mGy of baseline value
    2D kV (fluoroscopy mode) ±1cGy/min of baseline value
    All 3D imaging modes ±1cGy of baseline value
Image quality
    2D (spatial resolution, contrast) Baseline value
    3D (uniformity, spatial resolution, contrast)
Upgrade/Repair/Service
Verify / Reestablish QA baselines (as appropriate)
a

SRS=stereotactic radiosurgery; SBRT=stereotactic body radiation therapy.