Skip to main content
. 2014 Jan 6;15(1):187–201. doi: 10.1120/jacmp.v15i1.4507

Table 2.

Differences between the potentiometer and EPID header angles, including time delays and linear fits

EPID Image Headers
Uncorrected
(td=0)
Time Delay Corrected
(td=0.267s)
Boxcar Time Delay Corrected
Linac Trial Absolute Mean Diff. St. Dev.
%Within±1
%Within±1
%Within±1
#1 CW1 0.95 0.78 49.7 76.1 99.5
CW2 1.32 0.82 40.2 76.6 99.5
CW3 1.36 0.81 23.3 75.0 100
CCW1 1.06 0.89 41.2 73.8 98.9
CCW2 1.12 0.89 39.0 71.6 97.8
#2 CW1 1.10 0.74 35.6 80.6 100
CW2 1.01 0.75 46.4 81.8 98.9
CW3 1.18 0.73 37.0 79.2 100
CCW1 1.35 0.84 27.9 74.9 98.9
CCW2 1.06 0.87 43.4 73.6 99.4
CCW3 1.46 0.90 30.2 64.3 99.4
#3 CW1 1.65 0.78 21.6 75.8 98.9
CW2 1.18 0.79 40.3 74.3 99.5
CW3 1.51 0.78 25.8 77.4 100
CCW1 1.55 0.76 24.6 69.1 98.9
CCW2 1.24 0.73 34.0 80.1 100